Manufacturer: HTI (USA) 3-part differential automated hematology analyzer System Function Throughput: 60 samples per hour Parameters: (20) WBC, LYM#, MON#, GRA#, LYM%, MON%, GRA%, RBC, HGB, MCHC, MCH, MCV, RDW‐CV, RDW‐SD,HCT, PLT, MPV, PDW, PCT, PLCR and histograms for WBC, RBC and PLT Sample Volume: Prediluted 20μL Whole Blood 9.8μL Sample Processing: Open Tube or Cup, Manual Feed, Retractable Sample Probe Reagents: Lyse, Diluent, Cleanser, Cleaner Concentrate (maintenance)
Measuring Principles WBC Measurement:
Method: Impedance
Aperture (Ruby): 80um
Count Time: 2 (or 3) x 5 sec
Dilution Ratio: 1:269
3 Part Differentiation:
Method: Impedance + specific Lyse
Hgb Measurement:
Method: Photometry
Wavelength: 550nm
Count Time: 2 (or 3) x 5 sec
RBC/MCV/PLT Measurement:
Method: Impedance
Aperture (Ruby): 80um
Count Time: 2 (or 3) x 5 sec
Dilution Ratio: 1:17906
HCT, MCH, MCHC, RDW:
Calculated
Software Display: 10.4" LCD, Color touchscreen Operating System: Windows Language: English, Spanish, Russian, French, Ukrainian Memory: 250,000 Design: User Friendly, Rapid Access to Menus, Virtual Keyboard, User Profile (Access Level restrictions), Result Data Review Calibration: Automated, Manual Quality Control: 20 QC Files x 500 results; calculation of mean, SD, CV% per parameter, QC database, automatic charts, L-J of 500 results, X-B - 2000 QC data, X-R - 60 group (max 10 number of samples in each group) Connectivity: RS-232, USB x2, RJ-45 (Ethernet), LPT, PS/2 x2, LIS (output only), external printer capable, external mouse, keyboard & barcode reader Printer: Built-in thermal printer, 55mm paper Alerts: Real time messages, automatic maintenance and optional alerts, reagent level monitoring
Working Condition Power Supply: 110 ~ 240V, 50/60Hz, 96VA Temperature: 15°~ 35°C Humidity: 10 ~ 90% Dimension: 43.6cm x 36.7cm x 36.3cm (W x H x D) Weight: 18kg